ANALOG INPUT SECTION 11
Page 12-2
EXTERNAL TRIGGER
External trigger (CPU signal STADC) at J1-16 causes an
A-D conversion on a positive or rising pulse. Tie this
line to ground if you do not use this signal.
This line can be tied to the optional RTC output at J3-4.
Program mable, per iodic acquisitions from 2 to about
32,000 times/ second ar e possible. Execution time is
saved because conversions are started outside of a
program.
See application program AIN220-3. C for programming
information.
NOISE
An input channel can appear to be noisy (change
readings at rando m) if unuse d inputs are allowed to float.
To minimize noise (and increase accur acy), connect all
unused inputs to ground.
A high impedance input is, by definition, sensitive to
voltage pickup. Noise is minimized by running wires
away from AC power lines. A low impedance voltage
source helps to reduce noise pick up. Shielded cable can
help reduce noise from high impedance sources. Make
sure the shield is not used for power ground. Using the
shield for power ground defeats its purpose. Wire pairs
can also be twisted. 5-6 twists/foot provides a good
amount of noise cancellation.
Noise is defined in this section as any random change
from a known input. The amount of noise you can
exp ect und er nor ma l op er atin g circ um stance s is ±3
counts for any input range.
One way to compensate for noise is to take a number of
samples and average the results. Taking 7 or more
samples does, in theory, cancel out any effects of noise.
A problem with this is noise tends to group together.
Taking 7 readings at one time might show no change
from the norm. Another 7 readings might be all high.
If possible, try to spread out readings over a period of
time.
Noise is, by definition, random. If you wer e to plot out
the deviations from a norm, it would roughly resemble a
bell shaped curve. Exper iments on the RPC-220 have
shown that 99% of the readings are within the ±3 count
reading. N oise read ings wer e made with all inputs
shorted to ground.